Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12955/1585
Título : Calibration and validation of the Mini-fiber EC image analysis instrument mean fiber diameter through direct or primary measurements
Autor : Frank, Eduardo
Hick, E.N.
Castillo, M.V.H
Gaut, M.C
Mamani Cato, Rubén Herberht
Fecha de publicación : 13-dic-2021
Resumen : The Minifiber EC (MFEC) is a portable instrument for measuring the diameter of animal fibers. Its accuracy and precision have been estimated but by comparing its measurements with those of laboratory devices that had been calibrated on other devices in turn, not on a direct or primary measure of diameter. This work attempts to test direct measurements by gravimetry, Vernier mini caliper, microscope and the classic microprojector, using a non-deformable, high resistance synthetic fiber (Kevlar) for direct measurement. The MFEC instrument is calibrated with each mean fiber diameter obtained in direct measurements and its results are compared. The conclusions drawn are that it is possible to calibrate the MFEC instrument with direct measurements on Kevlar and measurement accuracy or tolerance of 0.28 microns is obtained. This indicates a very low biased mean fiber diameter measurement by MFEC.
Palabras clave : Minifiber EC
MFEC
Diameter of animal fibers
Editorial : MedCrave
URI : https://hdl.handle.net/20.500.12955/1585
metadata.dc.identifier.doi: https://doi.org/10.15406/jteft.2021.07.00290
metadata.dc.subject.ocde: https://purl.org/pe-repo/ocde/ford#4.04.02
Aparece en las colecciones: Artículos científicos

Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
Frank-et-al_2021_MEFC_Instrument.pdf647,62 kBAdobe PDFVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons